warning
In 2 minutes you will be logged out due to inactivity.
For information regarding change log and version number methodology, please click
here.
FW version 1.2.1 focuses on addressing a bug impacting the sensor table and improving reliability.
Description |
Impact |
Recommendation |
Bugs
- Revises sampling logic to improve reliability
- Adds sensor communication debug information to ping packets
|
Low impact |
No testing needed |
FW version 1.2.0 focused on optimizing battery backup power consumption. Multiple bug fixes have been included in this version as well.
Description |
Impact |
Recommendation |
Bugs
- Resolved an issue where DWE2 received an invalid MAC during the initial ethernet connection.
- Resolved a bug where logger lost serial number and token during operation.
- Resolved a bug where logger lost MAC address after power cycling.
- Resolved an issue where incorrect sensor model and serial number were sent after swapping to a single-channel sensor.
- Resolved an issue where audible alerts were not triggering due to configuration packet value rounding.
- Optimized battery backup power consumption.
|
Low impact |
No testing needed |
THIS DWE2 FIRMWARE IS A REQUIRED UPDATE - This firmware includes one bug fix which addresses the sensor configuration looping. This is a critical and important update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
Description |
Impact |
Recommendation |
Bugs
- Implemented a 10-minute delay between disconnecting a sensor and sending an empty
sensor packet to DicksonOne.
|
Low impact |
No testing needed |
Initial product release
Description |
Impact |
Recommendation |
New product release
|
Major impact |
Testing recommended, including possible regression testing |
Description |
Impact |
Recommendation |
- Updated RTC fallback time to 1/8/2019 midnight GMT
- Added functionality to update RTC fallback time to time of most recent message from D1
- Fixed issue where replaceable sensors with different sensor types couldn’t properly replace one another.
- Updated firmware will show new firmware version without need for reboot
- Fixed issue where config packet was sent on battery power
|
No impact |
No testing needed |
- Added rounding functionality to ensure display readings match what is shown in the DicksonOne interface
|
Minor impact |
Check new functionality |
Firmware version 1.9.8 is an identical release of 1.9.6 (since pulled from production servers) with the addition of the security certificate necessary to continue communicating beyond the phase out of a specific security certificate by Google.
Description |
Impact |
Recommendation |
Security
- Includes new SSL Certificate for continued communication between device and DicksonOne Servers.
|
No impact |
No testing needed |
Bug Fixes
- Fixed a bug where additional channel(s) was being created based on sensor type for RTMP sensors
- Reduced the size of ping packets to reduce watch dog timeouts caused by the ping packet
- Pairing/registration code is now the highest priority and will be displayed over all other messages (e.g. errors)
- Fixed multiple bugs allowing for devices to be properly configured reset at factory
- Fixed bug related to large numbers of sensors being connected over time
- Removed legacy code/functionality where ping packets were being sent every 6 hours when device was on battery
- Fixed issue with readings not being sent due to processing constraint
- Fixed “No AC” error message to appear whenever the device runs on battery
|
No impact |
No testing needed |
Improved Network Communication
- ARP cache cleared automatically after 20 minutes to allow device to connect to new access points
- Improved WiFi roaming to allow device to jump between different access points
|
No impact |
No testing needed |
Configuration
- Added support for reporting connection status to new Network Connection App
|
No impact |
No testing needed |
Device Exports
- Improved performance of USB file export; short button press now exports all data since last export and long press exports all data
|
Low impact |
No testing needed |
This update is required for continued operation of your DWEs after October 12, 2018.
Description |
Impact |
Recommendation |
This is the same as the pervious version (1.4.4) but has the new security certificates to address Google’s decision to distrust some existing certificates.
|
No impact |
No testing needed |
Description |
Impact |
Recommendation |
CRITICAL UPDATE
-
Adjusted packet intervals to use increased rate before registration and for 1 hour after being registered.
-
Fixed issue in wrapping functionality which can cause block chain errors.
-
Implemented patch routine to fix errors.
-
Fixed bug causing packet intervals to not update to slower rate correctly.
-
Fixed issue with readings having increased decimal places when sent to web.
-
Fixed bug causing min/max clear time not being saved correctly.
-
Further optimized power consumption when on battery.
-
Changed network layer to prioritize sending readings.
-
Added additional logging to ping packet.
-
Increased watchdog timeout for network layer.
|
No impact |
No testing needed |
Description |
Impact |
Recommendation |
-
Dropped support for SSL version less than TLS 1.2
-
Changed No AC message to only be shown when running on internal battery.
-
Updated thresholds corresponding to battery segment levels on LCD.
-
Added SSID to configuration packets.
-
Fixed issue with channels not cycling when on battery power.
-
Added indication of critical battery level when running on external battery.
|
No impact |
No testing needed |
Description |
Impact |
Recommendation |
- Disabled support for all SSL versions lower than TLS 1.2 to match website.
|
No impact |
No testing needed |